Case serial: SCH-2024-01-01
Title: Silicon Levin tube, 14FR
See the video for how to select your answer.
Patient Condition
- A 45-year-old female patient was inserted with a 14Fr Levin tube in the intensive care unit. As the patient was thin, the physician selected a 14Fr tube rather than a 16Fr tube.
- To check that the Levin tube was located inside the stomach, the physician put air in using a syringe and subsequently observed with a stethoscope.
- To double-check that the Levin tube was properly placed in the patient's stomach, a chest X-ray was taken.
- The radiologist failed to find the Levin tube from the patient's chest X-ray image. Refer to Figure (a).
- Normally, the Levine tube would be observed in the chest X-ray image as shown in Figure (b).
- Removed the Levin tube for patient safety and inserted another 14Fr tube.
- The patient retook a chest X-ray to check for the right placement of the Levin tube.
- The patient expressed discomfort in the nose area due to repeated insertion.
- Quality manager of the manufacturer detected that the radio-opaque line impregnated was narrower in width and contained smaller amount of barium sulfate at the tip, compared to the Levin tube of the same lot.
- The manufacturer decided to increase the width of the radio-opaque line to 8 mm for the Levin tube.
Events
Post-event management & Health effect
Investigations (Cause, Improvements)
Based on the above case, search appropriate IMDRF Adverse Event Terminology codes below by clicking on “Search” Button. To delete the code, click the code and then click the ‘Delete’ button. | |||
|
A | Medical Device Problem |
|
B | Cause Investigation – Type of Investigation |
|
|
C | Cause Investigation – Investigation Findings |
|
|
D | Cause Investigation – Investigation Conclusion |
|
|
E | Health Effects – Clinical Signs and Symptoms or Conditions |
|
|
F | Health Effects – Health Impact |
|
|
G | Medical Device Component |
|
|